20 C.F.R. Part 718, Appendix A

PART 718-STANDARDS FOR DETERMINING COAL MINERS' TOTAL DISABILITY OR DEATH DUE TO PNEUMOCONIOSIS

To return to the Table of Contents for 20 C.F.R. Part 718, click HERE .

APPENDIX B TO PART 718-STANDARDS FOR ADMINISTRATION AND INTERPRETATION OF PULMONARY FUNCTION TESTS. TABLES B1, B2, B3, B4, B5, B6

The following standards are established in accordance with section 402(f)(1)(D) of the Act. They were developed in consultation with the National Institute for Occupational Safety and Health (NIOSH). These standards are promulgated for the guidance of physicians and medical technicians to insure that uniform procedures are used in administering and interpreting ventilatory function tests and that the best available medical evidence will be submitted in support of a claim for black lung benefits. If it is established that one or more standards have not been met, the claims adjudicator may consider such fact in determining the evidentiary weight to be given to the results of the ventilatory function tests.

(1) Instruments to be used for the administration of pulmonary function tests shall be approved by NIOSH and shall conform to the following criteria:

(i) The instrument shall be accurate within 50 ml or within 3 percent of reading, whichever is greater.
(ii) The instrument shall be capable of measuring vital capacity from 0 to 7 liters BTPS.
(iii) The instrument shall have a low inertia and offer low resistance to airflow such that the resistance to airflow at 12 liters per second must be less than 1.5 cm H20/liter/sec.
(iv) The zero time point for the purpose of timing the FEV, shall be determined by extrapolating the steepest portion of the volume-time curve back to the maximal inspiration volume or by an equivalent method.
(v) Instruments incorporating measurements of airflow to determine volume shall conform to the same volume accuracy stated in subparagraph (1)(i) of this Appendix B when presented with flow rates from at least 0 to 12 liters per second.
(vi) The instrument or user of the instrument must have a means of correcting volumes to body temperature saturated with water vapor (BTPS) under conditions of varying ambient spirometer temperatures and barometric pressures.
(vii) The instrument used shall provide a tracing of either now versus volume or volume versus time during the entire forced expiration and volume versus time during the MVV maneuver. A tracing is necessary to determine whether the patient has performed the test properly. The tracing must be of sufficient size that hand measurements may be made within the requirement of subparagraph (1)(i) of this Appendix B. If a paper record is made it must have a paper speed of at least 2 cm/sec and a volume sensitivity of at least 10.0 mm of chart per liter of volume. The recorder tracing must display the entire FVC maneuver at a constant speed for at least 10 seconds after the onset of exhalation. This constant speed must be reached prior to the onset of exhalation.
(viii) The instrument shall be capable of accumulating volume for a minimum of 10 seconds after the onset of exhalation.
(ix) The forced expiratory volume in 1 sec (FEV1) measurement shall comply with the accuracy requirements stated in subparagraph (1)(i) of this Appendix B. That is, they shall be accurately measured to within 50 ml or with 3 percent of reading, whichever is greater.
(x) The instrument must be capable of being calibrated in the field with respect to the FEV1. This calibration of the FEV, may be done either directly or indirectly through volume and time base measurements. The volume calibration source shall provide a volume displacement of at least 3 liters and shall be accurate to within 30 ml.
(xi) For measuring maximum voluntary ventilation (MVV) the instrument shall have a response which is flat within 10 percent up to 4 Hz at flow rates up to 12 liters per second over the volume range. The time for exhaled volume integration or recording shall be no less than 12 sec. and no more than 15 sec. The indicated time shall be accurate to within 3 percent.

A recording of the spirometer tracing is required, and the volume sensitivity shall be such that 10 mm or more deflection corresponds to 1 liter volume.

(2) The administration of pulmonary function tests shall conform to the following criteria:

(i) Tests shall not be performed during or soon after an acute respiratory illness.
(ii) For the FEV1 and FVC, use of a nose clip is required. The procedures shall be explained in simple terms to the patient who shall be instructed to loosen any tight clothing and stand in front of the apparatus. The subject may sit, or stand, but care should be taken on repeat testing that the same position be used. Particular attention shall be given to insure that the chin is slightly elevated with the neck slightly extended. The patient shall be instructed to make a full inspiration, either from the spirometer or the open atmosphere, using a normal breathing pattern and then blow into the apparatus, without interruption, as hard, fast, and completely as possible. At least three forced expirations shall be carried out. During the maneuvers, the patient shall be observed for compliance with instructions. The expirations shall be checked visually for reproducibility from the now-volume or volume-time tracings. The effort shall be judged unacceptable when the patient:

(A) Has not reached full inspiration preceding the forced expiration; or

(B) Has not used maximal effort during the entire forced expiration; or

(C) Has not continued the expiration for least 5 sec. or until an obvious plateau in the volume-time curve has occurred; or

(D) Has coughed or closed his glottis; or

(E) Has an obstructed mouthpiece or a leak around the mouthpiece (obstruction due to tongue being placed in front of mouthpiece, false teeth falling in front of mouthpiece, etc.); or

(F) Has an unsatisfactory start of expiration, one characterized by excessive hesi-tation (or false starts), and therefore not allowing back extrapolation of time 0 (extrapolated volume on the volume-time tracing must be less than 10 percent of the FVC); or (G) Has an excessive variability between the three acceptable curves. The variation between the two largest FEV1's of the three acceptable tracings should not exceed 5 percent of the largest FEV1 or 100 ml, whichever is greater.

(iii) For the MVV, the subject shall be instructed before beginning the test that he or she will be asked to breathe as deeply and as rapidly as possible for approximately 15 seconds. The test shall be performed with the subject in the standing position, if possible. Care shall be taken on repeat testing that the same position be used. The subject shall breathe normally Into the mouthpiece of the apparatus for 10 to 15 seconds to become accustomed to the system. The subject shall then be instructed to breathe as deeply and as rapidly as possible, and shall be contin ually encouraged during the remainder 0 the maneuver. Subject shall continue the maneuver for 15 seconds. At least 5 minutes of rest shall be allowed between maneuvers. At least three MVV's shall be carried out. (But see §718.103(b).) During the maneuvers the patient shall be observed for compliance with instructions. The effort shall be judged unacceptable when the patient:

(A) Has not maintained consistent effort for at least 12 to 15 seconds; or

(B) Has coughed or closed his glottis; or

(C) Has an obstructed mouthpiece or a leak around the mouthpiece (obstruction due to tongue being placed in front of mouthpiece, false teeth falling in front of mouthpiece, etc.); or

(D) Has an excessive variability between the three acceptable curves. The variation between the two largest MVV's of the three satisfactory tracings shall not exceed 10 per-cent.

(iv) A calibration check shall be performed on the instrument each day before use, using a volume source of at least three liters, accurate to within 1 percent of full scale. The room air in the syringe is introduced into the spirometer once with a flow rate of approximately 0.5 liters per second (six seconds emptying time with a 3-liter syringe) and once with a higher flow rate of approximately 3.0 liters per second (one second emptying time with a 3-liter syringe). The volume measured by the spirometer shall be between 2.90 and 3.10 liters for both trials. Accuracy of the time measurement used in determining the FEVI shall be checked using the manufacturer's stated procedure and shall be within 3 percent of actual. The procedure described herein shall be performed as well as any other procedures suggested by the manufacturer of the spirometer being used.

(v)(A) The first step in evaluating a spirogram for the FEV1 shall be to determine whether or not the patient has performed the test properly or as described in (2)(ii) above. From the three satisfactory tracings, the forced expiratory volume in one second (FEV1) shall be measured and recorded. The largest observed FEV, shall be used in the analysis, corrected to BPTS.

(B) Only MVV maneuvers which demonstrate consistent effort for at least 12 seconds shall be considered acceptable. The largest accumulated volume for a 12 second period corrected to BPTS and multiplied by five is to be reported as the MVV.

[Editor's note: Pulmonary function study tables are not reproduced here.]