20 C.F.R. Part 718, Appendix A
PART 718-STANDARDS FOR DETERMINING COAL MINERS' TOTAL DISABILITY OR DEATH DUE TO PNEUMOCONIOSIS
To return to the Table of Contents for 20 C.F.R. Part 718, click HERE .
APPENDIX A TO PART 718-STANDARDS FOR ADMINISTRATION AND INTERPRETATION OF CHEST ROENTGENOGRAPHS (X- RAYS)
The following standards are established in accordance with sections 402(f) (1) (D) and 413(b) of the Act. They were developed in consultation with the National Institute for Occupational Safety and Health. These standards are promulgated for the guidance of physicians and medical technicians to insure that uniform procedures are used in administering and interpreting X-rays and that the best available medical evidence will be submitted in connection with a claim for black lung benefits. If it is established that one or more standards have not been met, the claims adjudicator may consider such fact in determining the evidentiary weight to be assigned to the physician's report of an X-ray.
(1) Every chest roentgenogram shall be a single posteroanterior projection at full inspiration on a 14 by 17 Inch film. Additional chest films or views shall be obtained if they are necessary for clarification and classification. The film and cassette shall be capable of being positioned both vertically and horizontally so that the chest roentgenogram will include both apices and costophrenic angles. If a miner is too large to permit the above requirements, then a projection with minimum loss of costophrenic angle shall be made.
(2) Miners shall be disrobed from the waist up at the time the roentgenogram is given. The facility shall provide a dressing area and, for those miners who wish to use one, the facility shall provide a clean gown. Facilities shall be heated to a comfortable temperature.
(3) Roentgenograms shall be made only with a diagnostic X- ray machine having a rotating anode tube with a maximum of a 2 mm source (focal spot).
(4) Except as provided in paragraph (5), roentgenograms shall be made with units having generators which comply with the following: (a) the generators of existing roentgenographic units acquired by the examining facility prior to July 27, 1973, shall have a minimum rating of 200 mA at 100 kVp; (b) generators of units acquired subsequent to that date shall have a minimum rating of 300 mA at 125 kVp.
NOTE: A generator with a rating of 150 kVp is recommended.
(5) Roentgenograms made with batterypowered mobile or portable equipment shall be made with units having a minimum rat- ing of lG6 mA at 110 kVp at 500 Hz, or 200 mA at 110 kVp at 60 Hz.
(6) Capacitor discharge, and field emission units may be used.
(7) Roentgenograms shall be given only with equipment having a beam-limiting device which does not cause large unexposed boundaries. The use of such a device shall be discernible from an examination of the roentgenogram.
(8) To insure high quality chest roentgenograms:
(i) The maximum exposure time shall not exceed 1/20 of a
second except that with single phase units with a rating less
than 300 mA at 125 kVp and subjects with chest over 28 cm
postero-anterior, the exposure may be increased to not more than
1/10 of a second;
(ii) The source or focal spot to film distance shall be at
least 6 feet;
(iii) Only medium-speed film and mediumspeed
intensifying screens shall be used;
(iv) Film-screen contact shall be maintained and verified at
6-month or shorter Intervals;
(v) Intensifying screens shall be inspected at least once a
month and cleaned when necessary by the method recommended by the
manufacturer;
(vi) All intensifying screens in a cassette shall be of the
same type and made by the same manufacturer;
(vii) When using over 90 kV, a suitable grid or other
means of reducing scattered radiation shall be used;
(viii) The geometry of the radiographic system shall
insure that the central axis (ray) of the primary beam is
perpendicular to the plane of the film surface and impinges on
the center of the film.
(9) Radiographic processing:
(i) Either automatic or manual film processing is
acceptable. A constant time-temperature technique shall be
meticulously employed for manual processing.
(ii) If mineral or other impurities in the processing water
introduce difficulty in obtaining a high-quality roentgenogram, a
suitable filter or purification system shall be used.
(10) Before the miner is advised that the examination is concluded, the roentgenogram shall be processed and inspected and accepted for quality by the physician, or if the physician is not available, acceptance may be made by the radiologic technologist. In a case of a substandard roentgenogram, another shall be made immediately.
(11) An electric power supply shall be used which complies with the voltage, current, and regulation specified by the manufacturer of the machine.
(12) A densitometric test object may be required on each roentgenogram for an objective evaluation of film quality at the discretion of the Department of Labor.
(13) Each roentgenogram made hereunder shall be permanently and legibly marked with the name and address of the facility at which it is made, the miner's DOL claim number, the date of the roentgenogram, and left and right side of film. No other identifying markings shall be recorded on the roentgenogram.